Laboratory freezer Snaige Medical LF60MD-P8000B0
For frozen storage of research samples, reagents, medicines and other laboratory materials at −35…−20 °C. Gross volume: 585 L; net volume: 500 L. Storage conditions and temperature for each product or material must follow its manufacturer’s instructions.
TEMPERATURE−35…−20 °CFactory setpoint −20 °C
VOLUME585 / 500 LGross / net
DOORSSolidDoor type
ENERGY CONSUMPTION2.234 kWhPer 24 hours
EXTERNAL DIMENSIONS705 × 886 × 2085 mmWidth × depth × height
INTERNAL DIMENSIONS535 × 712 × 1512 mmWidth × depth × height
Key features
- Control system: Electronic.
- Cooling system: Fan-assisted convection.
- Visual alarm: Yes.
- Audible alarm: Yes.
- Defrosting: Automatic hot gas.
- Temperature logger: Built-in; transfer via Bluetooth.
- Controller and alarm backup power duration: At least 24 h.
- Total number of shelves: 5 pcs.
- Maximum shelf load: 40 kg.
- The backup battery supports the controller, recording and alarms; it does not power cooling.
Technical specifications
| Volume | |
|---|---|
| Gross volume | 585 L |
| Net volume | 500 L |
| Dimensions and weight | |
| External dimensions (W × D × H) | 705 × 886 × 2085 mm |
| Internal dimensions (W × D × H) | 535 × 712 × 1512 mm |
| Net weight | 130 kg |
| Controls and electronics | |
| Control system | Electronic |
| Actual temperature display | Yes |
| User setpoint adjustment | Yes, via controller menu |
| Minimum and maximum temperature viewing | On display |
| Visual alarm | Yes |
| Audible alarm | Yes |
| High temperature alarm | Yes |
| Low temperature alarm | Yes |
| Door open alarm | Yes |
| Power failure alarm | Immediately after power failure |
| Controller and alarm backup power duration | At least 24 h |
| Functions on backup power | Controller, recording and alarms; no cooling |
| Temperature logger | Built-in; transfer via Bluetooth |
| Wireless connectivity | Bluetooth |
| Log application | CONTROLLA |
| Temperature recording interval | 20 min |
| Temperature log retention | 6 months |
| Number of recent alarm records | 5 records |
| Full alarm log overwrite | FIFO: oldest record is replaced by the new one |
| Alarm record data | Date, duration and highest temperature |
| Alarm indication retention | Until manual reset |
| Potential-free alarm contact | Yes |
| External alarm connection | Provided |
| Digital interface | RS485 |
| External sensor cable entry port | Yes |
| External sensor port diameter | 12.5 mm |
| Calibration block | No |
| Low temperature protection | Yes |
| Temperature specifications | |
| Minimum storage temperature | -35 °C |
| Maximum storage temperature | -20 °C |
| Factory temperature setpoint | −20 °C |
| Temperature fluctuation | 1.5 °C |
| Temperature gradient | 3.0 °C |
| Cooling | |
| Cooling system | Fan-assisted convection |
| Fan operation with compressor | Yes |
| Fan stops when door opens | Yes |
| Defrosting | Automatic hot gas |
| Refrigerant | R290 |
| Electrical and energy specifications | |
| Rated voltage | 220–240 V |
| Rated frequency | 50 Hz |
| Daily energy consumption | 2.234 kWh/24 h |
| Annual energy consumption | 816 kWh/year |
| Rated current | 2.3 A |
| Power input | 264 W |
| Mains cable length | 300 cm |
| Mains plug type | Euro plug |
| Cabinet, doors and feet | |
| Side wall material | White-coated steel |
| Interior material | AISI 304 stainless steel |
| Door type | Solid, white-coated steel |
| Door lock | Mechanical |
| Automatic door closing | Yes |
| Door hold-open angle | 90 ° |
| Reversible door | Provided |
| Removable door gasket | Yes |
| Interior lighting | Top horizontal LED lighting |
| Lighting activation | When door opens |
| Number of adjustable front feet | 2 pcs |
| Number of adjustable rear feet | 2 pcs |
| Minimum adjustable foot height | 97 mm |
| Maximum adjustable foot height | 159 mm |
| Number of lock keys | 2 pcs |
| Operating conditions | |
| Minimum ambient temperature | +10 °C |
| Maximum ambient temperature | +35 °C |
| Climate class | 7 |
| Maximum relative humidity | 75 % |
| Noise level | 53 dB(A) |
| Chamber equipment | |
| Total number of shelves | 5 pcs |
| Number of adjustable shelves | 5 pcs |
| Shelf width | 530 mm |
| Shelf depth | 650 mm |
| Maximum shelf load | 40 kg |
| Shelf material | Perforated stainless steel |
| Shelf mounting | On guide ribs of interior lining |
| Options ordered separately | |
| Additional shelves | Separate order |
| Branding | Separate order |
| Product identification | |
| Equipment type | Laboratory freezer |
| Model | LF60MD-P8000B0 |
| Factory code | LF60MD-P8000B01XXXXX2SN4B |
| EAN code | 4770104160028 |
| Packaging and product codes | |
| Packaging dimensions without pallet (W × D × H) | 768 × 948 × 2205 mm |
| Gross weight without pallet | 147 kg |
Application and limitations
Do not use for blood, plasma, other infusion fluids, chemically unstable, flammable or alkaline materials. Follow the storage conditions for the specific materials and the manufacturer’s manual.