Quantity
|
Total
|
||
|
|
|||
|
|
|||
Antech LamiFlow LF-V4 is a vertical laminar flow cabinet that protects non-hazardous samples, materials and equipment from particles. Its 1400 × 620 × 600 mm work zone, H14 HEPA filter and controlled 0.25–0.40 m/s airflow provide an ISO Class 5 environment. Available to order; price on request.
AIRFLOWVertical
WORKING WIDTH1400 mm
CLEANLINESSISO Class 5
Key benefits
- The H14 HEPA filter provides at least 99.995% efficiency at 0.3 µm and is supported by a washable pre-filter.
- A DC ECM fan with automatic compensation maintains the selected airflow velocity.
- The 4.3-inch touchscreen displays velocity, pressure, temperature, humidity and filter life.
- Visual and audible alarms indicate airflow deviations and filter-life limits.
- Cabinet and UV timers, operating-hour counters, password protection and an RS485 port simplify monitoring.
- The SUS304 work tray, glass side walls and four castors support cleaning and relocation.
Technical specifications
| Model | LF-V4 |
|---|---|
| MPN | 3121141 |
| Manufacturer | Qingdao Antech Scientific Co., Ltd., China |
| Type | Laminar flow cabinet for product protection |
| Airflow direction | Vertical |
| Air velocity | 0.25–0.40 m/s, adjustable |
| Nominal size | 4 ft / 1400 mm |
| External dimensions with stand (W × D × H) | 1500 × 720 × 1780 mm |
| External dimensions without stand (W × D × H) | 1500 × 720 × 1095 mm |
| Work zone (W × D × H) | 1400 × 620 × 600 mm |
| Usable work area | 0,87 m² |
| Working sash height | 190-330 mm |
| Maximum opening | 520 mm |
| Control | 4.3-inch touchscreen |
| Monitoring | Velocity, differential pressure, temperature, humidity, UV and filter hours |
| Timers | Cabinet and UV lamp |
| Alarms | Velocity deviation, sash position, filter life |
| Interface | RS485 |
| Main filter | H14 HEPA, ≥99.995% at 0.3 µm |
| Pre-filter | Washable |
| Cleanliness class | ISO Class 5 |
| Fan | DC ECM |
| Work tray | SUS304, 1.5 mm thick |
| Side walls | 6 mm glass |
| Front sash | 6 mm tempered, UV-resistant glass |
| Power sockets | 2; 1320 W / 6 A total |
| Power supply | 220 V, 50/60 Hz |
| Rated power / current | 450 W / 2 A |
| LED lighting | 14 W × 1 |
| UV | 30 W × 1 |
| Illuminance | 700 lux |
| Noise level | 55 dB(A) |
| Net weight | 175 kg |
| Stand / castors | Steel stand; 4 lockable castors |
| Packing dimensions (W × D × H) | 1610 × 820 × 1970 mm |
| Gross weight | 235 kg |
Applications and limitations
- For media preparation, non-hazardous culture work, pharmaceutical, optical, electronic and other operations that require a clean product-protection airflow.
- This is not a biological safety cabinet: LamiFlow does not protect the operator or environment from hazardous bioaerosols.
- Do not use with toxic, volatile, radioactive or pathogenic materials without a separate risk assessment and suitable containment equipment.
- Do not obstruct the work zone or air channels; final positioning and commissioning must account for room ventilation.
- Factory options, delivery time and final price are confirmed in the quotation.