Quantity
|
Total
|
||
|
|
|||
|
|
|||
LaboGene ScanLaf Atlas Pro Compact Class I Safety Cabinet is a narrow benchtop cabinet for local operator and environmental protection when handling powders, aerosols and chemicals. The series occupies approximately 630 mm of bench width.
Important: this is a Class I cabinet and does not protect the product, because unfiltered room air enters the work zone. Sterile or contamination-sensitive samples require a suitable Class II cabinet.
Atlas Pro configuration
Air passes through the HEPA filter and a final carbon filter before returning to the laboratory. The carbon type must be selected for the substances and vapours involved. An internal extract fan maintains controlled airflow.
Atlas Pro should be selected only after assessing the chemicals and carbon-filter service life. Processes requiring guaranteed external discharge need the Atlas Duct configuration.
Key features
- stable 0.4–0.7 m/s inflow through a fixed 400 mm opening;
- 99.999% HEPA efficiency at 0.3 µm;
- eye-level LCD, automatic start/stop, timer and lighting control;
- time and airflow alarms;
- front service access and a compact benchtop footprint.
Confirmed specifications
| Protection class | Class I — operator and environmental protection; no product protection |
| HEPA filter | 110 mm depth; 99.999% efficiency at 0.3 µm |
| Inflow velocity | 0.4–0.7 m/s |
| Work opening | Fixed 400 mm height |
| External dimensions | Approx. 627 × 340 × 1075 mm (W × D × H), based on the manufacturer drawing |
| Working area | Approx. 573 × 285 × 535 mm (W × D × H), based on the manufacturer drawing |
| Bench width occupied | Approx. 630 mm |
| Fan | Energy-efficient EC fan; current consumption below 0.2 A |
| Controls | LCD panel, automatic start/stop, time and lighting control, airflow alarms |
| Service access | Front access to filters, PCB, sensors and adjustments |
Options
PIR sensor, 700–1000 mm electrically adjustable stand, USB port, airflow sensors, pressostat and external ventilation signals. The final configuration is defined after assessing the process and substances.