Quantity
|
Total
|
||
|
|
|||
|
|
|||
EMSAS ECI-3 is a 1078 L platelet incubator operating at +22 to +24 °C, with a +22 °C factory setpoint. It accommodates 6 EAJ-05/09 or 3 EAJ-L09 agitators. Agitators are selected and supplied separately.
Key benefits
- Password-protected microprocessor control maintains the selected temperature.
- Thirty-day data memory supports USB export.
- Opening the door pauses the fan and agitation; operation resumes after closure.
- Monitoring covers temperature, open door, power, sensor and agitator-stop conditions.
- A stainless-steel chamber and fan-assisted circulation support organised loading.
- Double-glazed doors have magnetic seals and locks.
Technical specifications
| Model | EMSAS ECI-3 |
| Type | Platelet incubator |
| Operating range | +22…+24 °C |
| Factory setpoint | +22 °C |
| Volume | 1078 L |
| External dimensions (W × D × H) | 697 × 1215 × 2000 mm |
| Gross weight | 200 kg |
| Polyurethane insulation | 50 mm |
| Agitator compatibility | 6 EAJ-05/09 or 3 EAJ-L09 agitators |
| Internal chamber | Cr-Ni 304 stainless steel |
| External cabinet | Powder-coated galvanised steel |
| Air circulation | Fan-assisted, multiple air streams |
| Controller | Microprocessor |
| Temperature sensor | NTC |
| Data recording | 30-day memory; USB export |
| Door | Double glazing, magnetic seal and lock |
| Internal lighting | Yes |
| Castors | Two with brakes and two without |
| Power | 230 W |
| Electrical supply | 220 V / 50 Hz or 110 V / 60 Hz; confirm configuration |
| Refrigerant | R134a / R600 / R290; confirm configuration |
| Options | Thermal printer; SMS and email functions |
Applications and configuration
For a controlled platelet-storage environment in blood banks, transfusion departments and laboratories. Validate the actual load, bag layout, compatible agitator and local operating procedure.
The base supply does not include a confirmed agitator. The 0.1 °C measurement figure is not a claim of spatial chamber uniformity. Backup power supports the control panel and recording, not refrigeration. Confirm printer, messaging options, voltage, refrigerant, warranty and final configuration in the quotation.