Medical pharmaceutical refrigerator Snaige Medical MD14DD-P600QX2
For storage of vaccines, medicines and other pharmaceutical products at +2…+8 °C. Gross volume: 130 L; net volume: 100 L. Storage conditions and temperature for each product or material must follow its manufacturer’s instructions.
TEMPERATURE+2…+8 °CFactory setpoint +5 °C
VOLUME130 / 100 LGross / net
DOORSSolidDoor type
ENERGY CONSUMPTION0.482 kWhPer 24 hours
EXTERNAL DIMENSIONS560 × 620 × 870 mmWidth × depth × height
INTERNAL DIMENSIONS465 × 392 × 725 mmWidth × depth × height
Key features
- Control type: Electronic.
- Cooling system: Forced air circulation by fan.
- Visual alarm: Yes.
- Audible alarm: Yes.
- Defrosting: Automatic.
- Temperature recording: Built-in controller memory.
- Controller and alarm backup power duration: At least 24 h.
- Total number of shelves: 4 pcs.
- Maximum load per shelf: 20 kg.
- The backup battery supports the controller, recording and alarms; it does not power cooling.
Technical specifications
| Volume | |
|---|---|
| Gross volume | 130 L |
| Net volume | 100 L |
| Dimensions and weight | |
| External dimensions (W × D × H) | 560 × 620 × 870 mm |
| Internal dimensions (W × D × H) | 465 × 392 × 725 mm |
| Net weight | 38 kg |
| Gross weight without pallet | 41 kg |
| Controls and electronics | |
| Control type | Electronic |
| Temperature display | Actual air temperature near evaporator |
| Service temperature setpoint adjustment | By manufacturer or official representative |
| Visual alarm | Yes |
| Audible alarm | Yes |
| Fault code display | Yes |
| Door not closed alarm | Yes (code dor) |
| Low temperature protection | Yes |
| Potential-free alarm contact | Provided |
| Additional external sensor port | Provided |
| External sensor port diameter | 12.5 mm |
| Interior lighting type | Top LED strip |
| Lighting activation | When door opens; continuous operation available |
| Calibration block | Provided |
| Temperature recording | Built-in controller memory |
| Temperature recording interval | 20 min |
| Temperature log retention | 6 months |
| Number of stored alarm records | 5 records |
| Wireless connectivity | Bluetooth |
| Data application | CAREL CONTROLLA |
| Temperature and alarm log export | Via CONTROLLA application |
| Minimum temperature display | Yes |
| Maximum temperature display | Yes |
| Mains power failure alarm | Immediately after power failure |
| Controller and alarm backup power duration | At least 24 h |
| Cooling from backup battery | Not provided |
| External alarm connection | Provided |
| Alarm indication until manual reset | Yes |
| Digital interface | RS485 |
| Local network connection | Via external system and RS485 |
| Independent protective temperature controller | Provided |
| Low temperature protection threshold | +2 °C |
| Temperature specifications | |
| Minimum storage temperature | +2 °C |
| Maximum storage temperature | +8 °C |
| Factory temperature setpoint | +5 °C |
| User setpoint adjustment | Not provided |
| Temperature fluctuation | 0.5 °C |
| Temperature gradient | 1.8 °C |
| Positive deviation from setpoint | +0.5 °C |
| Negative deviation from setpoint | −1.4 °C |
| Cool-down time from +25 to +5 °C¹ | 40 min |
| Temperature rise time from +5 to +10 °C¹ | 40 min |
| Recovery time after door opened for 1 min¹ | 8 min |
| Cooling | |
| Cooling system | Forced air circulation by fan |
| Defrosting | Automatic |
| Manual defrost activation | Via controller menu |
| Refrigerant | R600a |
| Protection against contact with cold surfaces | Evaporator protective cover |
| Fan stops when door opens | Yes |
| Electrical and energy specifications | |
| Rated voltage | 220–240 V |
| Rated frequency | 50 Hz |
| Rated current | 0.5 A |
| Rated power input | 90 W |
| Energy consumption per 24 hours | 0.482 kWh |
| Annual energy consumption | 176 kWh |
| Heat emission | 122 kJ/h |
| Sound power level relative to 1 pW | 42 dB(A) |
| Power cable length | 240 cm |
| Plug type specified by manufacturer | Euro plug |
| Cabinet, doors and feet | |
| Exterior material | Steel |
| Exterior colour | White |
| Interior material | HIPS polystyrene |
| Door type | Solid |
| Self-closing door | Yes |
| Reversible door | Provided |
| Door lock type | Mechanical |
| Number of keys supplied | 2 pcs |
| Door gasket replacement | Provided |
| Number of adjustable front feet | 2 pcs |
| Number of rear feet | 2 pcs |
| Operating conditions | |
| Climate class | 7 |
| Minimum ambient temperature | +10 °C |
| Maximum ambient temperature | +35 °C |
| Maximum relative humidity | 75 % |
| Stacking two appliances | Declared by manufacturer for this model |
| Chamber equipment | |
| Total number of shelves | 4 pcs |
| Number of adjustable shelves | 2 pcs |
| Number of bottom shelves | 2 pcs |
| Adjustable shelf width | 420 mm |
| Adjustable shelf depth | 324 mm |
| Bottom shelf 1 width | 465 mm |
| Bottom shelf 1 depth | 263 mm |
| Bottom shelf 2 width | 445 mm |
| Bottom shelf 2 depth | 180 mm |
| Maximum load per shelf | 20 kg |
| Shelf material | Wire grid |
| Shelf coating | White polymer paint |
| Options ordered separately | |
| Perforated metal shelves | Separate order |
| Additional shelf | Separate order |
| Branding | Separate order |
| SNAIGĖ CONNECT GSM notification system | Separate order |
| Organisers | Separate order |
| Product identification | |
| Product name | Medical pharmaceutical refrigerator |
| Model | MD14DD-P600QX2 |
| Brand | Snaige |
| Series | MAX PLUS |
| Manufacturer | AB SNAIGĖ |
| Country of manufacture | Lithuania |
| Manufacturer address | Pramonės g. 6, LT-62175 Alytus, Lithuania |
| Packaging and product codes | |
| Factory code without pallet | MD14DD-P600QX23AXXXZ1SN4B |
| EAN without pallet | 4770104014031 |
| Packed dimensions without pallet (W × D × H) | 570 × 640 × 950 mm |
| Factory code with pallet | MD14DD-P600QX23AXXXZ1SN4A |
| EAN with pallet | 4770104014031 |
| Gross weight with pallet | 49.5 kg |
| Packed dimensions on pallet (W × D × H) | 570 × 710 × 1080 mm |
¹ Time specifications apply to an empty chamber at an ambient temperature of +25 °C. Recovery is measured after the door is opened for 1 minute. Temperature rise time is not a guaranteed product storage period during a power outage.
Application and limitations
Do not use for blood, plasma, other infusion fluids, chemically unstable, flammable or alkaline materials. Follow the storage conditions for the specific materials and the manufacturer’s manual.